21 CFR PART 820 · ISO 13485:2016 Published Aug. 20, 2026 · Updated Oct. 21, 2026

What the QMSR Adds to ISO 13485:2016

21 CFR Part 820 has required a quality system that complies with ISO 13485:2016 since February 2, 2026. The part adds a short set of FDA-specific requirements.

Shelves of labeled procedure binders and folders in a document archive.

21 CFR 820.7 incorporates ISO 13485:2016 by reference. Under 21 CFR 820.10(a), a manufacturer must document a quality management system that complies with the applicable requirements of the standard. Under 21 CFR 820.10(e), failure to comply with an applicable requirement makes a device adulterated under section 501(h) of the Federal Food, Drug, and Cosmetic Act.

Part 820What it addsISO clause
21 CFR 820.10(b)(1)A system to assign unique device identification under part 830.7.5.8
21 CFR 820.10(b)(2)Traceability procedures under part 821, if applicable.7.5.9.1
21 CFR 820.10(b)(3)Notification to the FDA of complaints that meet the reporting criteria of part 803.8.2.3
21 CFR 820.10(b)(4)Advisory notices handled under part 806.7.2.3, 8.2.3, 8.3.3
21 CFR 820.10(c)Design and development, clause 7.3, required for class II, class III, and listed class I devices.7.3
21 CFR 820.10(d)Traceability for implantable devices (clause 7.5.9.2), required for devices that support or sustain life when failure under proper use can reasonably be expected to cause significant injury.7.5.9.2
21 CFR 820.35Required content for complaint, servicing, and unique device identifier records.4.2.5, 8.2.2, 7.5.4
21 CFR 820.45Procedures for labeling and packaging integrity, with an accuracy examination before release.7.5.1

Part 820 now has six sections with text: 820.1, 820.3, 820.7, 820.10, 820.35, and 820.45. Sections 820.20 through 820.30 are reserved. A procedure, training record, or audit checklist that cites the older numbering, such as 21 CFR 820.100 for corrective and preventive action, cites a section that no longer exists.

Sixteen of the 27 device warning letters issued on or after February 2, 2026 and posted through October 20, 2026 allege a quality-system violation. Eleven of the sixteen cite a section of Part 820. Six cite ISO 13485 clauses. One of those also cites a section. Two letters cite no section of the CFR at all. One of them is the Linemaster letter.

Which of your procedures still cite a section of Part 820 that no longer exists?

Sources: 21 CFR Part 820 on the eCFR (read October 21, 2026) and the FDA’s warning letters issued on or after February 2, 2026 and posted through October 20, 2026.